SR 1000 Specifications

 

Input

Measurement points:
1,2,3 (pen) and 6 (dot) point

Input signals, Measurement range
and Measurement Range Limits:
Input
RANGE
Measurement Range
DC voltage (V)
20 mV
-20.00 to 20.00 mV
60 mV
-60.00 to 60.00 mV
200 mV
-200.0 to 200.0 mV
2 V
-2.000 to 2.000 mV
6 V
-6.000 to 6.000 mV
20 V
-20.00 to 20.00 V
Thermocouples (TC)
RANGE
Measurement Range (°C)
Measurement Range (°F)
R * 1
0.0 to 1760.0 °C
32 to 3200 °F
S * 1
0.0 to 1760.0 °C
32 to 3200 °F
B * 1
0.0 to 1820.0 °C
32 to 3308 °F
K * 1
-200.0 to 1370.0 °C
-328 to 2498 °F
E * 1
-200.0 to 800.0 °C
-3280 to 1472.0 °F
J * 1
-200.0 to 1100.0 °C
-3280 to 2012.0 °F
T * 1
-200.0 to 400.0 °C
-3280 to 752.0 °F
N * 1
0.0 to 1300.0 °C
32 to 2372 °F
W * 2
0.0 to 2315.0 °C
32 to 4199°F
L * 3
-200.0 to 900.0 °C
-3280 to 1652.0 °F
U * 3
-200.0 to 400.0 °C
-3280 to 752.0 °F
RTD
RANGE
Measurement Range (°C)
Measurement Range (°F)
Pt100*4
-200.0 to 600.0 °C
-3280 to 1112.0 °F
Pt100*4
-200.0 to 550.0 °C
-3280 to 1022.0 °F
Contact input (operation recording)
RANGE
Measurement Limit
Dl 1
voltage input
Less than 2, 4V; OFF;
2,4 V or more: ON (TTL)
Dl 2
contact input
Contact ON/OFF
*1 R, S, B, K, E, J, T, N:IEC 584-1-1995, DIN IEC 584, JIS C 1602-1995
*2 W:W-5% Rd/W-26% Rd (Hoskins Mfg. Co.) ASTM E988
*3 L: Fe-CuNi, DIN 43710
*3 U: Cu-CuNi, DIN 43710
*4 Pt100: JIS C 1604-1997, IEC 751-1995, DIN IEC 751-1996
*4 JPt100: JIS C 1604-1989, JIS C 1606-1989
 
DCA: DC Current Signal Input
(using external resistor)
Measurement and Recording Accuracy:
Input Type
RANGE
Measurement (digital display)
Recording (analog)
Measurement accuracy
Maximum resolution
Recording accuracy
DC voltage
(DC V)
20mV
+/-(0.2% of rdg +3 digits)
10uV

Measurement accuracy +/- (0.3% of recording span)

60mV
+/-(0.2% of rdg +2 digits)
10uV
200mV
+/-(0.2% of rdg +2 digits)
100uV
2V
+/-(0.1% of rdg +2 digits)
1mV
6V
+/-(0.3% of rdg +3 digits)
1mV
20V
+/-(0.3% of rdg +2 digits)
10mV
Thermocouple (TC)
R

+/-(0.15% of rdg +°C)
Except R,S: 0to100°C,+/-3.7°C;
100 to 300°C, +/-1.5°C
B: 400 to 600°C, +/-2°C
No grantee under 400°C

0.1°C
Measurement accuracy +/- (0.3% of recording span)
S
B
K
+/-(0.15% of rdg +0.7°C)
Except at -200 to -100°C,
+/-(0.15% of rdg +1°C)
E
+/-(0.15% of rdg +0.5°C)
J
+/-(0.15% of rdg +0.5°C)
Except at -200 to -100°C,
+/-(0.15% of rdg +0.7°C
0.1°C
T
N
+/-(0.15% of rdg +0.7°C)
W
+/-(0.15% of rdg +1°C)
0.1°C
L
+/-(0.15% of rdg +0.5°C)
Except at -200 to -100°C,
+/-(0.15% of rdg +0.7°C)
0.1°C
U
RTD
Pt100
+/-(0.15% of rdg +0.3°C)
0.1°C
Measurement accuracy +/- (0.3% of recording span)
JPt100

For more information regarding Marlin Instruments,
Please call us at 1-800-321-6648 or email us at marlincle@marlinmfg.com

 

 
Go to top of page

Marlin Manufacturing Corporation · 12800 Corporate Drive · Parma · Ohio · 44130

Phone: (216) 676-1340 or (800) 321-6648·
Fax: (216) 676-1344 · Email: marlincle@marlinmfg.com
Company Profile Contact Us Home Home

Copyright 2003 Marlin Mfg. Corporation. All Rights Reserved.

Site Design: tlhalfacre@ameritech.net & Abstract Productions